| Features | |
|---|---|
Features  | 
    a) Steam Temp. 121.4 oC at 15 psi.  | 
  
Specification  | 
    As per BS 2646-5:1993. Autoclaves for sterilization in    laboratories Methods of Test for function and performance.   | 
  
Temp Controller  | 
    PID (Optional: Combined Temp. Pressure & water level Controller)  | 
  
Pressure Controller  | 
    Mechanically operated pressure/steam release valve.  | 
  
Pressure Gauge  | 
    Analog Dial type Pressure Gauge (Optional: Digital Pressure Meter)  | 
  
Temperature Sensor  | 
    Standard coupling type PT 100 Sensor  | 
  
Heat Exchanger  | 
    The electrical heat exchanger transfer latent heat to the clean fill water to generated steam under pressure.  | 
  
Water Level Control  | 
    Low water level cut off electronic circuit protects the electrical heat exchanger from low water level inside the chamber, this Cut's off the power supply to the heat exchanger.  | 
  
Steam Outlet  | 
    Steam generated from the unit is raised and exits thought the electrically operated steam Purging solenoid valve from the outlet  | 
  
| Models | |||||
|---|---|---|---|---|---|
Capacity  | 
    35 ltrs  | 
    52 ltrs  | 
    119 ltrs  | 
    178 ltrs  | 
    213 ltrs  | 
  
Inner Size  | 
    30 dia X 50 h  | 
    35 dia X 55 h  | 
    40 dia X 75 h  | 
    55 dia X 75 h  | 
    55 dia X 90 h  | 
  
Inner Chamber  | 
    Stainless Steel 304 quality  | 
  ||||
External  | 
    Stainless Steel 304 quality  | 
  ||||
Range  | 
    Ambient to 15 psi pressure / 121.40C  | 
  ||||
Controller  | 
    PID (Optional: Combined Temp. Pressure & water level Controller)  | 
  ||||
Sensor  | 
    PT 100  | 
  ||||
Accuracy  | 
    +/-1.oC  | 
  ||||
Uniformity  | 
    +/-0.5oC  | 
  ||||
Max. Ramp up time  | 
    45 min. up to 121.40C without load with respect to ambient  | 
  ||||
Insulation  | 
    Air Bank 50 mm  | 
  ||||
Top Lid  | 
    With Dome Shape Die press Stainless Steel Plate  | 
  ||||
Optional  | 
  |||||
Printer Output  | 
    Printer output port RS232  | 
  ||||
Validation Facility  | 
    Data logging facility for temperature validation report.  | 
  ||||
Capacity (Watts)  | 
    2000  | 
    3000  | 
    4000  | 
    5000  | 
    6000  | 
  
Power Consumption  | 
    10 amps  | 
    15 amps  | 
    20 amps  | 
    25 amps  | 
    30 amps  | 
  
Net Weight approx.  | 
    15kgs  | 
    20kgs  | 
    30kgs  | 
    50kgs  | 
    55kgs  | 
  
Electric requirements AC230 V & 50Hz cycle.  | 
  |||||