Features | |
---|---|
Features |
a) Steam Temp. 121.4 oC at 15 psi. |
Specification |
As per BS 2646-5:1993. Autoclaves for sterilization in laboratories Methods of Test for function and performance. |
Temp Controller |
PID (Optional: Combined Temp. Pressure & water level Controller) |
Pressure Controller |
Mechanically operated pressure/steam release valve. |
Pressure Gauge |
Analog Dial type Pressure Gauge (Optional: Digital Pressure Meter) |
Temperature Sensor |
Standard coupling type PT 100 Sensor |
Heat Exchanger |
The electrical heat exchanger transfer latent heat to the clean fill water to generated steam under pressure. |
Water Level Control |
Low water level cut off electronic circuit protects the electrical heat exchanger from low water level inside the chamber, this Cut's off the power supply to the heat exchanger. |
Steam Outlet |
Steam generated from the unit is raised and exits thought the electrically operated steam Purging solenoid valve from the outlet |
Models | |||||
---|---|---|---|---|---|
Capacity |
35 ltrs |
52 ltrs |
119 ltrs |
178 ltrs |
213 ltrs |
Inner Size |
30 dia X 50 h |
35 dia X 55 h |
40 dia X 75 h |
55 dia X 75 h |
55 dia X 90 h |
Inner Chamber |
Stainless Steel 304 quality |
||||
External |
Stainless Steel 304 quality |
||||
Range |
Ambient to 15 psi pressure / 121.40C |
||||
Controller |
PID (Optional: Combined Temp. Pressure & water level Controller) |
||||
Sensor |
PT 100 |
||||
Accuracy |
+/-1.oC |
||||
Uniformity |
+/-0.5oC |
||||
Max. Ramp up time |
45 min. up to 121.40C without load with respect to ambient |
||||
Insulation |
Air Bank 50 mm |
||||
Top Lid |
With Dome Shape Die press Stainless Steel Plate |
||||
Optional |
|||||
Printer Output |
Printer output port RS232 |
||||
Validation Facility |
Data logging facility for temperature validation report. |
||||
Capacity (Watts) |
2000 |
3000 |
4000 |
5000 |
6000 |
Power Consumption |
10 amps |
15 amps |
20 amps |
25 amps |
30 amps |
Net Weight approx. |
15kgs |
20kgs |
30kgs |
50kgs |
55kgs |
Electric requirements AC230 V & 50Hz cycle. |